Alarm level
ALM LEV
Options: HIGH, MED (medium) and LOW.
More than one option can be selected at one time.
Alarm type
ALM TYPE
Options: TECH. (technical) and PHYS. (physiological).
Both options can be selected at one time.
Trigger Conditions
A nurse call signal will be triggered only if all the following conditions are met:
1.
The nurse call function is enabled.
2.
An alarm of the preset alarm level and alarm type comes out.
3.
The monitor is not in the Alarms Paused or the System Silenced status.
NOTE
If no option in ALM LEV or ALM TYPE is selected, the nurse call signal will not
be triggered in whatever condition.
The nurse call function can’t be used as the main alarm notice method. Medical
staff must combine the audible and visual alarms, the clinical vital signs of the
patient to determine the patient’s situation.
In the Alarms Paused or the System Silenced status, the nurse call function of the
monitor will be disabled automatically.
4.7.5 CO2 User Maintain
Selecting CO2 USER MAINTAIN >> in USER MAINTAIN menu opens the CO2 USER
MAINTAIN menu. The options contained in this menu are relative with the CO2 module that
you monitor is equipped with. For details, please refer to chapter CO2 Monitoring.
4.7.6 Gas Calibrate
Selecting GAS CALIBRATE >> in USER MAINTAIN menu opens the GAS CALIBRATE
menu. The options contained in this menu are relative with the AG module that you monitor
is equipped with. For details, please refer to chapter Anesthesia Gas Monitoring.
4-25